Step 3. Max. Capacity Setting ( display)
Input Max. Capacity of Scale with No. keys.
Input Capacity and press ENTER KEY, and move to next step.
※ Caution
(Max. capacity value / division value) can not be over 20,000.(as Indicator resolution is 1/200,00).
Step 4. Zero Balance setting ( display)
Make empty the scale part, and press ENTER KEY.
Indicator check the current Zero balance and save the value and move next step.
Step 5. Span value calculation ( )
Input prepared Test weight value with No. keys.
And press ENTER KEY.
Then, display will show and then, load prepared test weight unit on the scale.
After a few seconds(to remove the vibration effect), press ENTER KEY.
Then, indicator will calculate Span value and move the next step.
※ Caution
For the precise Span calibration, please prepare Test weight unit, at least 10% of Max. capacity
of Scale.
Step 6. Check Span Value and Finish the Calibration( )
Check the Calculated Span value.
And after 3sec, will displayed automatically and move to weighing Mode.