30 / 44
Other Setting
X.X.X.X.X.X.
(0.0.0.0.1.0)
You can set “EMPTY” Range.
Within set range, indicator will not display current weight and just
display “Zero”.
“0.000” setting : When Net Zero, “Zero” status lamp and Near Zero
relay will be output.
“0.190” setting : Within 190, “Zero” Status lamp and Near Zero relay
will be output.
Within this “Zero Range setting”, all the weight value will be
displayed, As “0”
Zero Value Deduction Setting
Display value with deduction, as much as set value.
Ex.)Set 1000, actual weight 3000, then display 2000, only.
Analogue Output Setting (only for the analogue option installation)
At the set weight value, analogue output will be maximized.
Ex.) Set 5000, then weight 5000 20mA or 10V will be output
At this function, you can check the Calculated Span value.
※ If you have difficulty to process Calibration again, the best way to
matching the net weight and display weight is doing Calibration
process once again.
Check Current DATE data or you can Change to new date
Check Current TIME data or you can Change to new TIME