6. InterFace
1. Serial Interface
RS-232C Serial Interface is sensitive/weak for electric Noise.
So, please isolate with AC power cable and use shield cable to reduce the electric noise effect.
1-1. Signal Format
. Type : EIA-RS-232C
. Communication Method : Half-Duplex, Full Duplex, Asynchronous
. Serial Baud Rate : Selectable on “F-function31”
. Data Bit : 8(No Parity mode, only)Bit Refer “F30”.
. Stop Bit : 1
. Parity Bit : Non, Even, Odd (Selectable on “F-function 30”) - Refer “F30”
. Code : ASCII
STX 02H
ETX 03H
CR 0DH
LF 0AH
1
+9V
LSB
0
1
2
3
4
5
MSB
6
0
-9V
Start Bit
Data Bit
Parity
Bit
Stop
Bit
1.2. Standard Data Format(1) : ID Number will not be transferred. (Refer “F-function 35”)
,
,
+/-
Including Decimal
point
k
g
CR
LF
Header
1.
Header
2.
Data Byte ( 8 or 9 byte )
(F38 setting)
Unit
. Header 1. : OL : Over Load, Under Load
ST : Display weight “Steady”
US : Display “Un-Steady”
. Header 2. : NT : Net-Weight
GS : Net-Weight, under TARE
. Data Bit(Number) 2B(H) : “+” Plus
2D(H) : “-“ Minus