52
5. Once the sample pads have been tared, add the
correct amount of sample to the pad and spread the
product using the proper sample spread technique.
The sample bar representing sample weight will
change from BLUE to GREEN when it is within the
designated Sample Weight range created during
method development
*If you are unsure how to add the correct amount of
sample, or the correct sample spread technique,
please refer to the videos titled “How to Run a
Sample” for assistance. If you still have questions,
please contact CEM Corporation Process Application
Support (800) 726-3331
6. After the sample has been added and properly
spread, again center the pads on the balance pan,
gently close the SMART 6 cover, and press the START
button.
The system will automatically record a stable initial
weight, and the SMART 6 will then begin its heating
procedure based on the parameters for the method
that was chosen. If the system does not change
screens within 30 seconds of pressing start, press