ECO 6 Digital Color Doppler Ultrasound System
115
Customer Service department or Agents of manufacturer.
If no image or menu but image, please check following malfunction table. If the malfunction can’t be solved,
please contact with Sale Office, Customer Service department or Agents of manufacturer.
8.4 Malfunction Check
Serial
number
Malfunction
Reason
Measures
1
Switch button lights but
power LED not
Battery lose efficacy, Adapter
works irregular
Check the connector
between cable and power
2
Power Led lights but LED
no image
The interval time is too short to
restart
Restart after 1 minute
3
LED display character menu
but no scan image
Launch power, gain or STC
control errors
Not connect to probe or the
probe connection is not correct
Device is in freezing condition
Control launch power , gain
or STC control
Make sure of right
connection
Exit from freezing condition
4
Abnormal image
Exam mode errors
Image processing setting errors
Whether Exam mode is
proper or not
Adjust image processing
setting or set it as default
5
Probe works improperly
1.The plug plugs loosely
2.Internal circuit protects
1.Extract the probe and
reinsert
2.Restart
6
No OB calculation package
menu
Do not select the OB application
before scanning.
Select the OB application
7
PRINT-key doesn’t work
1.The connected printer is
approved
2.Printer power is not on
3.Printer is not connect well
1.Change the approved
printer
2.Turn on the printer
3.Connect the printer again