Programmable AC Source 61501/61502/61503/61504 User’s Manual
8-20
[SOURce :] PULSe : FREQuency
Description : This command sets the frequency during the duty cycle of
PULSE mode.
Query Syntax : [SOURce :] PULSE : FREQuency?
Parameters : <NR2>, valid range: 15.00 ~ 1000.00 (unit: Hz)
Return Parameters : <NR2>
[SOURce :] PULSe : SHAPe
Description : This command selects the waveform buffer for PULSE mode.
Query Syntax : [SOURce :] PULSE : SHAPe?
Parameters : A | B
Return Parameters : A | B
[SOURce :] PULSe : SPHase
Description : This command sets the start phase angle of duty cycle of
PULSE mode.
Query Syntax : [SOURce :] PULSE : SPHase?
Parameters : <NR2>, valid range: 0.0 ~ 359.9
Return Parameters : <NR2>
[SOURce :] PULSe : COUNt
Description : This command sets the number of times that the pulse is
executed before it is completed.
Query Syntax : [SOURce :] PULSE : COUNt?
Parameters : <NR2>, valid range: 0 ~ 65535
Return Parameters : <NR2>
[SOURce :] PULSe : DCYCle
Description : This command sets the duty cycle of PULSE mode.
Query Syntax : [SOURce :] PULSE : DCYCle?
Parameters : <NR2>, valid range: 0 % ~ 100 %
Return Parameters : <NR2>
[SOURce :] PULSe : PERiod
Description : This command sets the period of the PULSE mode.
Query Syntax : [SOURce :] PULSE : PERiod?
Parameters : <NR2>, valid range: 0 ~ 99999999.9 (unit: ms)
Return Parameters : <NR2>
OUTPut : MODE
Description : This command sets the operation mode
Query Syntax : OUTPut : MODE?
Parameters : FIXED | LIST | PULSE | STEP | SYNTH | INTERHAR
Return Parameters : FIXED | LIST | PULSE | STEP | SYNTH | INTERHAR
TRIG
Description : This command sets PULSE mode in OFF execution state after
setting OUTPut : MODE PULSE. If users want to change the
parameters, it’s necessary to set TRIG OFF then OUTPut :
MODE FIXED. Then, set OUTPut : MODE PULSE again to
get ready to set TRIG ON.
Query Syntax : TRIG : STATE?
Parameters : OFF | ON
Return Parameters : OFF | RUNNING