Programmable AC Source 61501/61502/61503/61504 User’s Manual
4-2
1. The Password of entering calibration procedure is " 7377 ", then
press
ENTER.
2. Before calibrating the AC source, users should read the procedure
in details. Or it may lose some data in memory because of improper
operation.
The display changes to CALIBRATION CHOICE PAGE as below after keying the right
password.
V OUT AND MEAS.: output voltage and voltage measurement accuracy calibration.
I MEAS.: current measurement accuracy calibration.
EXT Vref.: external Vref calibration.
Enter Password:
(You can get password in user's manual! )
Enter Password: * * * *
(You can get password in user's manual! )
CALIBRATION CHOICE = 1_ [ CALIB ]
1. V OUT AND MEAS. 2. I MEAS.
3. EXT Vref.
PAGE CHOICE = 4_
1. SETUP 2.CONF 3.OUTPUT 4. MANUAL CALI
5.
LIST 6. PULSE 7.
STEP
8. HAR 9.
SYN
10. INTERHAR