Programmable AC Source 61501/61502/61503/61504 User’s Manual
8-10
Query Syntax : FETCh [ : SCALar] : VOLTage : ACDC?
MEASure [ : SCALar] : VOLTage : ACDC?
Return Parameters : <NR2>
FETCh [ : SCALar] : VOLTage : DC?
MEASure [ : SCALar] : VOLTage : DC?
Description : These queries return the DC composition of output voltage
which is being output at the output terminals.
Query Syntax : FETCh [ : SCALar] : VOLTage : DC?
MEASure [ : SCALar] : VOLTage : DC?
Return Parameters : <NR2>
8.6.2.2 OUTPUT Sub-system
OUTPut
[: STATe]
: RELay
: SLEW
: VOLTage
: AC
: DC
: FREQuency
: COUPling
: IMPedance
: STATe
: RESistor
: INDuction
: MODE
: PROTection
: CLEar
OUTPut [: STATe]
Description :This command enables or disables the output of the AC source.
Disable output is to set an output voltage amplitude at 0 Volt.
Query Syntax : OUTPut [: STATe]?
Parameters : OFF | ON
Return Parameters : OFF | ON
OUTPut : RELay
Description : This command sets output relay on or off.
Query Syntax : OUTPut : RELay?
Parameters : OFF | ON, ON sets the output relay of the AC source on
(closed). OFF sets the output relay of the AC source off
(open).
Return Parameters : OFF | ON
OUTPut : SLEW : VOLTage : AC
Description : This command sets the slew rate of the AC output voltage.
Query Syntax : OUTPut : SLEW : VOLTage : AC?
Parameters : <NR2>, valid range: 0.000V/ms ~ 1200.000V/ms
Return Parameters : <NR2>