Programmable AC Source 61605 User’s Manual
4-2
4.2 MANUAL CALI Functional List
Users can choose " 4. MANUAL CALI " in CHOICE PAGE to enter the calibration procedure.
Before showing the calibration items, for safety reason, user must enter password. The
password is shown in this manual, in order to confirm the user read the manual before doing
the calibration procedure.
1. The Password of entering calibration procedure is " 7377 ", then
press ENTER .
2. Before calibrating the AC source, users should read the procedure in
details. Or it may lose some data in memory because of improper
The display changes to CALIBRATION CHOICE PAGE as below after keying the right
password.
V OUT AND MEAS.: output voltage and voltage measurement accuracy calibration.
I MEAS.: current measurement accuracy calibration.
EXT Vref.: external Vref calibration.
PAGE CHOICE = 4_
1. SETUP 2.CONF 3.OUTPUT 4. MANUAL CALI
Enter Password :
( You can get password in user's manual ! )
Enter Password : * * * *
( You can get password in user's manual ! )
CALIBRATION CHOICE = 1_ [ CALIB ]
1. V OUT AND MEAS. 2. I MEAS.
3. EXT Vref.