Programmable AC Source 61705 User’s Manual
4-2
1. The password to enter the calibration procedure is " 7377 ", then
press ENTER.
2. Before calibrating the AC Source, users should read the procedures
in detail. Or it may lose some data in memory because of improper
After entered the correct Password, the display shows PHASE CHOICE PAGE.
Choose the phase to calibrate, the display shows CALIBRATION CHOICE PAGE.
V OUT AND MEAS.: Output voltage and voltage measurement accuracy calibration.
I MEAS.: Current measurement accuracy calibration.
( You can get password in user's manual ! )
Enter Password : * * * *
( You can get password in user's manual ! )
CALIBRATION CHOICE = 1_
1. V OUT AND MEAS. 2. I MEAS.
PAGE CHOICE = 4_
1. SETUP 2.CONF 3.OUTPUT 4. MANUAL CALI
5.
LIST 6. PULSE 7.
STEP
8. INTERHAR
PAGE CHOICE = 4_
1. SETUP 2.CONF 3.OUTPUT 4. MANUAL CALI
5.
LIST 6. PULSE 7.
STEP
8. INTERHAR