Remote Operation
8-25
Parameters : <NR2>, valid range: 0.01 ~ 2400.00 (unit: Hz)
Return Parameters : <NR2>
[SOURce :] INTerharmonics : LEVel
Description : This command sets the r.m.s. magnitude of the sweep wave
in percentage of fundamental.
Query Syntax : [SOURce :] INTerharmonics : LEVel?
Parameters : <NR2>, valid range: 0% ~ 30% at 0.01 Hz ~ 500 Hz
0% ~ 20% at 500.01 Hz ~ 1000 Hz
0% ~ 10% at 1000.01 Hz ~ 2400 Hz
Return Parameters : <NR2>
[SOURce :] INTerharmonics : DWELl
Description : This command sets the dwell time of sweep wave.
Query Syntax : [SOURce :] INTerharmonics : DWELl?
Parameters : <NR2>, valid range: 0.01 ~ 9999.99 (unit: sec)
Return Parameters : <NR2>
[SOURce :] INTerharmonics : MODe
Description : This command sets the mode of interharmonic function.
Query Syntax : [SOURce :] INTerharmonics : MODe?
Parameters : ALL | #1 | #2 | #3
Return Parameters : ALL | #1 | #2 | #3
OUTPut : MODE
Description : This command sets the operation mode
Query Syntax : OUTPut : MODE?
Parameters : FIXED | LIST | PULSE | STEP | INTERHAR
Return Parameters : FIXED | LIST | PULSE | STEP | INTERHAR
TRIG
Description : This command sets INTERHARMONICS mode in OFF, ON,
PAUSE or CONTINUE execution state after setting OUTPut :
MODE INTERHAR. If users want to change the
parameters, it’s necessary to set TRIG OFF then OUTPut :
MODE FIXED. Then, set OUTPut : MODE INTERHAR
again to get ready to set TRIG ON.
Query Syntax : TRIG : STATE?
Parameters : OFF | ON | PAUSE | CONTINUE
Return Parameters : OFF | RUNNING | COMPLETE
FETCh [:SCALar] : INTerharmonics : FREQuency?
MEASure [:SCALar] : INTerharmonics : FREQuency?
Description : These queries return the sweeping frequency superimposed
on fundamental voltage.
Query Syntax : FETCh : INTerharmonics : FREQuency?
MEASure : INTerharmonics : FREQuency?
Return Parameters : <NR2>
8.6.2.13 STATUS Sub-system
STATus