Spyder 3 CL User's Manual 63
DALSA 03-032-20008-00
Digital Signal Processing
To optimize camera performance, digital signal processing should be completed after any
analog adjustments.
FPN Correction
Performing FPN Correction
Syntax: Performs FPN correction and eliminates FPN noise by subtracting
away individual pixel dark current.
Syntax:
ccf
Notes:
Perform all analog and digital adjustments before
performing FPN correction.
Perform FPN correction before PRNU correction.
Refer to
Calibrating the Camera to Remove Non-Uniformity
(Flat Field Correction)on page
60 for a procedural overview
on performing flat field correction.
To save FPN coefficients after calibration, use the
wfc
command. Refer to section
7.8 Saving and Restoring PRNU
and FPN Coefficients for details.
Related Commands:
ccp, wfc
Example:
ccf
Setting a Pixel’s FPN Coefficient
Purpose: Sets an individual pixel’s FPN coefficient.
Syntax
sfc x i
Syntax Elements:
x
The pixel number from 1 to sensor pixel count.
i
Coefficient value in a range from 0 to 2047 (12 bit LSB).
Example:
sfc 10 50