System 9700 User Manual
Page 43 of 56
3.3.5.4. SW limits
The limits of the “Arbitrary point method” ramp profile SW are:
The set value must be given in ppm units.
The time between two points can be between 0.2 to 65535 seconds.
Maximum numbers of stacks: = 15
Maximum number of lines in a stack: = 15
The limits of the “Equal time slot method” ramp profile SW are:
The set value must be given in a floating point representation normalized to 1.000000.
The time slot may be between 0.0125 to 1 second given in a floating point representation
normalized to 1.0000.
Maximum numbers of stacks = 1
Maximum number of time slots in a stack = 1000
The limits of the “Auto Slew Rate method” ramp profile SW are:
Maximum positive slew rate time = 10 sec,
Maximum negative slew rate time = 10 sec,
Minimum positive slew rate time = 5 ms,
Minimum negative slew rate time = 5 ms,
Number of points = 80 with a minimum interval of 1.25ms