LT-2320
22
SPECIFICATIONS
UNIT SPECIFICATIONS
DV6x CPAP Device Specications
Size 3.7” (9.4 cm) H x 6.1” (15.5 cm) W x 5.9” (15 cm) D
Weight 1.93 lbs (0.9 kg) CPAP only; 3.8 lbs. (1.74 kg) including DV6HH heated humidier
Electrical Requirements AC 100-240V~, 50/60 Hz
Electrical Requirements DC 11-17 VDC, 5.2 Amps
Maximum Power Consumption 65 watts max from AC power source (ow generator only)
Pressure Range 3-20 cmH
2
O
Operating Temperature Range 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range 15% to 93% RH non-condensing
Operating Atmospheric Conditions Range 1060 hPa - 700 hPa (~ 1,400 ft below sea level to 9,800 ft)
Storage & Transportation Temperature Range -13˚F to 158˚F (-25˚ C - +70˚ C)
Storage & Transportation Humidity Range 15% to 93% RH non-condensing
Maximum Limited Pressure 20 cmH
2
O under normal use, 30 cmH
2
O under single fault
Sound Pressure Level (tested per ISO 17510-1:2007) 26.6 dBA
Sound Power Level 34.6 dBA
Filter Specications
Standard Filter > 3.0 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Maximum Flowrate (per ISO 17510-1:2007)
Test Pressures 3.0 cmH
2
O 7.0 cmH
2
O 12.0 cmH
2
O 16.0 cmH
2
O 20.0 cmH
2
O
Measured pressure at the patient
connection port (cmH
2
O)
2.9 6.6 11.4 15.0 19.0
Average ow at the patient
connection port (l/min)
78.2 123.5 166.8 157.7 138.0
Long Term Static Pressure Accuracy ± 0.5 cmH
2
O
Dynamic Short-Term Pressure Accuracy with and without DV5HH Humidier (per ISO 17510-1:2007)
Waveform
Volume
[mL]
Breath Rate
[min-1]
Set Pressure
[cmH
2
O]
Pressure Accuracy
CPAP only
[Pk-Pk cmH
2
O]
Pressure Accuracy
CPAP with Humidier
[Pk-Pk cmH
2
O]
Sinusoidal cycle with I:E = 1:1 500 10, 15, 20
3.0 0.5 0.5
7.0 0.5 0.5
12.0 1.0 1.0
16.0 1.0 1.0
20.0 1.0 1.0
Additional Specications
Equipment classication with respect to protection from electric shock Class II
Degree of protection from electric shock Type BF Applied Part
Degree of protection against ingress of liquids
IP21 Ingress Protection protected against nger access to hazardous
parts and vertically falling water drops
Mode of operation Continuous
Equipment not suitable for use in the presence of a ammable anesthetic mixture with air, oxygen or nitrous oxide. See instructions for use with
supplemental oxygen.
US Patents
SmartCode
®
Technology is protected under US Patent 8649510
Approvals
Meets RTCA/DO 160 – DO-160 section 21 Category M for battery operation only Airline use.