GENERAL INFORMATION
DV6x CPAP devices will operate on 12V DC power that has been
inverted to 120V AC power. NOTE: See DC Power Chart below for
the correct inverter specications.
DC Power Chart for Vehicle & 12V Marine Battery
Conguration of
Use
Inverter
Specication for
use in vehicle
Battery
Specication for
stand-alone use
Battery Hours
@ 20 cmH
2
O
If using a vehicle battery for multiple nights without powering the vehicle’s
motor during use, power the vehicle during the day to replenish the battery.
CPAP device
with heated
humidication
setting of 5
120V AC Modied
or Quasi Sine
Wave Inverter with
200w capacity
120 amp-hour,
deep-cycle
Marine Type
8 hours
120V AC Modied
or Quasi Sine
Wave Inverter with
300w capacity
230V AC Modied
or Quasi Sine
Wave Inverter with
400w capacity
240 amp-hour,
deep-cycle
Marine Type
16 hours
CPAP device
without heated
humidication
80 amp-hour,
deep-cycle
Marine Type
8 hours
E. SUPPLEMENTAL OXYGEN
WARNING-OXYGEN USE
Oxygen supports combustion. To avoid possible physical injury,
do not smoke while using this device with supplemental oxygen.
Do not use this device near hot objects, volatile substances or
sources of open ames.
Always start the CPAP blower before starting the oxygen ow.
Stop the oxygen ow before stopping the CPAP blower.
Never allow the oxygen source to run continuously while
connected to the DeVilbiss CPAP device if the device is not in
use. If the DeVilbiss CPAP device is not being used, turn off the
oxygen ow.
CPAP pressure must be set higher than 8 cmH
2
O for
supplemental oxygen use to prevent patient breathing pattern
from forcing oxygen back into the CPAP.
At a xed ow rate of supplemental oxygen, the inhaled oxygen
concentration will vary depending on the pressure setting, patient
breathing pattern, mask selection and leak rate.
This warning applies to most types of CPAP devices.
CAUTION
Oxygen is a prescribed gas and should only be administered
under the supervision of a physician.
The ow setting of the oxygen source must be specied by a
physician.
Maximum oxygen pressure is 50 psi (345 kPa). Maximum oxygen
ow rate is 10 LPM.
If a physician prescribed supplemental oxygen, it can be added one of
two ways:
Attach oxygen tubing directly to the oxygen port on the mask
Insert the optional oxygen adapter into the device’s air supply
port. Then, insert the air supply tubing and the oxygen tubing to
the adapter.
Always START CPAP before starting oxygen ow.
Always STOP oxygen ow before stopping CPAP.
Oxygen Adapter
(7353D-601)
F. SETTING PRESSURES AND
FEATURES
Factory Default Settings
Factory default settings are installed on the IntelliPAP 2/DeVilbiss
BLUE DV6x CPAP device prior to shipping.
Setting Default Value
Operating Mode DV63 - CPAP
DV64 - AutoAdjust
CPAP pressure Default 10.0 (DV63 initial setting is 8.0,
default is 10.0)
Upper Pressure Limit (UPL) 15.0 for DV64 AutoAdjust mode
Lower Pressure Limit (LPL) 7.0 for DV64 AutoAdjust mode
Apnea % 90
Apnea Duration 10 seconds
Hypopnea % 40
Hypopnea Duration 10 seconds
Ramp Pressure 4.0 cmH
2
O
Ramp Time 20 minutes
Split Night - Inhibit Time 240 minutes
Number of Sessions 0 (off)
Usage Threshold 4 hours
Wakeup Alert 0 (Off)