MAINTENANCE AND FAULT FINDING
L025317 Issue 1 March 2011 5-27
(6) Insert the new main filter into the machine ensuring that the filter label
is in front and the airflow arrow is pointing up.
(7) Turn the filter locking handle 180° clockwise, so that it is horizontal
and in the filter locked position.
(8) Close the cabinet door, turning the catch fully clockwise.
Changing the Chemical Filter (DPX2000)
WARNING: Full protective clothing must be worn and
correct disposal methods used immediately.
(1) Turn off the extractor.
(2) Open the cabinet door.
(3) Turn the filter locking handle 180° anti-clockwise so that it is
horizontal and in the filter unlocked position.
WARNING: This filter is heavy and must be removed with
care.
(4) The filter will drop slightly to allow release. Pull the filter forward and
remove completely.
(5) Insert the new filter into the machine ensuring that the airflow arrow is
in front and the filter label is pointing up.
(6) Turn the filter locking handle 180° clockwise, so that it is horizontal
and in the filter locked position.
(7) Close and lock the cabinet door, turning the catch fully clockwise.
Changing the Chemical Pad (DPX2000)
WARNING: Full protective clothing must be worn and
correct disposal methods used immediately.
(1) Turn off the extractor
(2) Open the cabinet door.
(3) Remove the Super Filter Bag.
(4) Lift the Chemical Pad out of the bottom of the compartment and
dispose of using approved procedures.
(5) Place a new Chemical Pad onto the bottom of the compartment.
(6) If required, fit a new Filter Bag as on page 5-27.
(7) Close and lock the cabinet door, turning the catch fully clockwise.