Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 113
Tests
The automatic test consists of a pretest (the initialization), which may require
manual actions by the user, and a fully automatic main test.
On devices with an O
2 sensor, the system test checks whether an O2 calibration is
required. If more than 7 days have passed since the last calibration of the O
2
sensor, an O
2 calibration will be automatically performed.
On devices with an integrated patient-gas measurement module, automatic
calibrations are performed regularly during operation.
7.4.1.1 Execution
WARNING
Risk of patient injury
During the system test, the device is pressurized.
► To prevent patient injury, do not perform the system test and leakage test if a
patient is connected.
1. Touch the button for the system test.
2. Complete the tabular checklist (1).
It is possible to switch from the tabular checklist to the walk-through mode, if
required. To do this, touch the button (2). For further information see: "Checklist
in the walk-through mode", page 125.
3. If the O
2 sensor is to be calibrated despite a valid calibration, touch the
Calibrate button (3).