Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 113
Tests
The automatic test consists of a pretest (the initialization), which may require
manual actions by the user, and a fully automatic main test.
On devices with an O
2 sensor, the system test checks whether an O2 calibration is
required. If more than 7 days have passed since the last calibration of the O
2
sensor, an O
2 calibration will be automatically performed.
On devices with an integrated patient-gas measurement module, automatic
calibrations are performed regularly during operation.
7.4.1.1 Execution
WARNING
Risk of patient injury
During the system test, the device is pressurized.
To prevent patient injury, do not perform the system test and leakage test if a
patient is connected.
1. Touch the button for the system test.
2. Complete the tabular checklist (1).
It is possible to switch from the tabular checklist to the walk-through mode, if
required. To do this, touch the button (2). For further information see: "Checklist
in the walk-through mode", page 125.
3. If the O
2 sensor is to be calibrated despite a valid calibration, touch the
Calibrate button (3).
35072
1
2
54
3
System test - checklist

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals