Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 129
Tests
2. Open the flow control valve (1). Set the desired O2 flow.
The O
2 flowmeter indicates a flow, the breathing bag fills, and the inflow of gas is
audible.
3. Close the flow control valve (3).
4. Set the O
2 switch (4) back to the Aux. O2 position.
3511235111
3
4