130 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Tests
7.5.3 Breathing circuit
Prerequisite:
Breathing system is complete and locked.
Breathing system cover is fitted.
1. Hoses (1) and filters, e.g., at the Y-piece (2), are connected properly.
2. Extendable hoses (3) are extended to the length intended for use.
Do not change the length of the hoses after the test is done.
3. Remove the water from the hoses (4).
3511535117
2
1
3
4