Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 131
Tests
4. Press the O2+ key. The inflow of gas is audible. Keep the key pressed until the
pressure exceeds 15 hPa (cmH
2O).
5. Pull off the Y-piece and plug it on again. The pressure drops.
7.5.4 Valves
7.5.4.1 Checking the O2 flush
1. Press the O
2+ key until the pressure exceeds 15 hPa (cmH2O).