138 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Tests
6. After the calibration is complete, remove the sealing cap and plug it back in the
holder.
Insert the O
2 sensor (4) back into the sensor port and turn it clockwise until it
reaches the end position.
7. After the calibration, the fully automatic main test is continued.
Note on the expiry of the life span of the O
2 sensor cell
The life span of the O
2 sensor cell is 2 years and is monitored by the device. As
soon as there are 28 days or fewer of the life span left, the remaining life span will
be displayed in the system test. When the life span has expired, a corresponding
message and the following illustration will be displayed:
If the O
2 calibration is possible despite the expired O2 sensor cell, the O2
measurement will continue to be available. With an expired O
2 sensor cell, a yellow
test result (operational with limitations) can be achieved at best.
Replace the sensor cell when the O
2 sensor can no longer be calibrated, see
page 69.
3968039676
4

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals