162 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Operation
8.5.5 Alarm behavior during the maneuvers
During the maneuvers, some alarms are adapted as follows:
Some technical alarms stop the maneuver automatically, e.g., alarms caused by a
sensor failure.
Alarm
Maneuvers
Insp./Exp. hold
One-step
recruitment
Multi-step
recruitment
Man. insp./
Insp. hold
Exp. hold
Pressure alarms:
Airway pressure
high
--- ---
Alarm limit is
adjusted to
Pressure + 5
Alarm limit is adjusted
to Pinsp max + 5
Airway press.
continuously high
Stopped
1)
1) Existing alarms are preserved. No new alarms will be issued.
---
Alarm limit is
adjusted to
Pressure + 3
Alarm limit is adjusted
to half way between
PEEP max and Pinsp
max.
Example:
PEEP max = 20 hPa
Pinsp max = 40 hPa
Adjusted alarm limit =
30 hPa
Inspiratory pres-
sure not
achieved
Switched off Switched off Switched off ---
Volume alarms:
Minute volume
low
Is suppressed during the maneuver and for an additional
45 seconds after the end of the maneuver.
---
Apnea alarms:
Apnea (no pres-
sure)
Switched off
---
Apnea (no flow)
Apnea (no CO
2)
Stopped
1)

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals