Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 171
Operation
8.7.3 Customizing the current view
Waveforms and parameter fields can be customized as follows:
1. Touch the waveform or parameter field.
The Quick setup dialog opens.
WARNING
Risk due to inadequate monitoring
National and medical regulations may require certain parameters to be displayed.
Always consider the relevant regulations when configuring the screen layout.
2. For Content (1), select the desired content.
For a list of the possible screen content, see page 225.
3. For Scale (2), select the desired setting.
Unsaved changes to the current view are kept until the next device restart.
8.7.3.1 Restoring the current view
The changes to the current view can be canceled.
1. Open the Views dialog.
2. Touch the Restore current view button.
8.7.4 Using loops
Prerequisite: The device has the "advanced ventilation monitoring" or "loops and
trends" software option.
The following loops are available:
Pressure-Volume loop
Flow-Volume loop
35686
2
1
Quick setup

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals