Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 219
Configuration
11.5.6.4 Vertical tab "System test"
11.5.6.5 Vertical tab "Logbook"
Headline/
Parameter
Setting range
Description
General
Always use walk-
through mode
On; Off Sets the walk-through mode function.
When On is configured, system tests
and leakage tests will always be exe-
cuted in walk-through mode.
Verify O
2 delivery
and circuit
(requires sample
line)
On; Off Specifies whether the automatic tests
will check whether the O
2 supply actu-
ally delivers oxygen and whether the
breathing circuit is correctly connected.
Prerequisite: The sample line is con-
nected to the Y-piece or the filter on the
Y-piece during the tests.
If On is configured, the device automat-
ically checks the O
2 supply and the cor-
rect connection of the breathing circuit.
If Off is configured, the O
2 supply is not
checked. In order to verify the correct
connection of the breathing circuit,
additional manual test steps must be
completed in the checklist.
Test gas supply
Central O
2 supply On; Off Specifies which gas supplies are tested
during the automatic system test.
If Auto On is configured, only the gas
supplies are checked whose actual
availability is guaranteed after auto-
matic start-up.
At least one O
2 source must be tested
during the system test.
Central Air supply On; Off
Central N
2O sup-
ply
On; Off
O
2 cylinder
Air cylinder
N
2O cylinder
On; Off
Headline/
Parameter
Setting range
Description
A logbook entry with measured values is created Enables the creation of additional log-
book entries with measured values.
Every 1 min; 2 min; 5 min; 10 min; 15 min Generates periodic entries.
For all high-
priority alarms
On; Off Generates entries in the event of an
alarm.
For all medium-
priority alarms
On; Off