Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 275
Technical data
15.6 Fresh-gas delivery with a mechanically controlled gas
mixer and flow tubes
15.7 Ventilator
Resolution of the value displayed in the
status display
0.1 L/min
Total flow tube
Standardization All data are standardized to STPD conditions.
Range Off to 10 L/min
Accuracy ±10 % of the set value at 100 % O
2 and for flows
>1 L/min
Resolution of displayed value 0.5 L/min (in the range from 1 to 10 L/min)
Standardization All data are standardized to STPD conditions.
O
2 concentration 21 to 100 Vol%
Setting range for fresh-gas flow Off to at least 10.0 L/min (O
2, Air, and N2O)
Measurement of the fresh-gas flow for O
2
and N
2O
Range 0.02 to 10.0 L/min
Accuracy ±10 % of the delivered flow (from 0.1 to 10.0 L/min)
±50 % of the delivered flow (below 0.1 L/min)
Resolution of displayed value 0.05 L/min (from 0.05 to 1.00 L/min)
0.50 L/min (from 1.00 to 10.0 L/min)
Measurement of the fresh-gas flow for Air
Range 0.2 to 10.0 L/min
Accuracy ±10 % of the delivered flow (from 0.6 to 10.0 L/min)
±50 % of the delivered flow (below 0.6 L/min)
Resolution of displayed value 0.2 L/min (from 0.2 to 1.0 L/min)
0.5 L/min (from 1.0 to 6.0 L/min)
2.0 L/min (from 6.0 to 10.0 L/min)
Design Electronically driven ventilator, fresh-gas decoupled
Time-based settings
Respiratory rate RR
Setting range 3 to 100 /min
Increment value 1 /min
Accuracy ±10 % of the set value or ±1 /min (the greater value
applies)
Minimum respiratory rate RRmin in PSV mode
Setting range Off, 3 to 60 /min
Increment value 1 /min