Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 277
Technical data
Number of breaths at each level
(Breaths/Step), with the exception of the high-
est level in the Multi-step recruitment maneu-
ver
Setting range 1 to 20
Increment value 1
Number of breaths at the highest level
(Breaths@Max) in the Multi-step recruitment
maneuver
Setting range 1 to 20
Increment value 1
Reminder for repeated One-step recruitment
or Multi-step recruitment maneuver
Setting range Off, 10 to 180 min
Increment value 10 min
Accuracy ±1 min
Volume-based and flow-based settings
Tidal volume VT
Setting range 10 to 1500 mL
Setting range for "advanced neonatal
support" software option
5 to 1500 mL
Increment value 5 mL in the range from 10 to 100 mL
10 mL in the range from 100 to 1500 mL
Increment value for "advanced neonatal
support" software option
1 mL in the range from 5 to 25 mL
5 mL in the range from 25 to 100 mL
10 mL in the range from 100 to 1500 mL
Accuracy ±5 % of the set value or ±15 mL (the greater value
applies) in the range above 150 mL
±10 % of the set value or ±10 mL (the greater value
applies) in the range from 20 to 150 mL
±50 % of the set value or ±4 mL (the greater value
applies) in the range below 20 mL
Information The applied tidal volume is automatically adjusted to
compensate for the compliance of the breathing circuit.
As soon as CO
2 respiratory phases are detected, the
sample gas flow for the patient-gas measurement mod-
ule is additionally compensated.
With set values above 1400 mL, the specified accuracy
can be reduced if a very long breathing circuit or a
breathing circuit with a very high compliance is used.
Trigger threshold Trigger
Setting range 0.3 to 15.0 L/min
Increment value 0.1 L/min in the range from 0.3 to 2.0 L/min
1 L/min in the range from 2.0 to 15.0 L/min
Accuracy ±20 % of the set value or ±1 L/min (the greater value
applies)