278 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Technical data
Peak inspiratory flow 180 L/min or higher
Results from the following set values:
- VT and Ti in volume-controlled ventilation modes
- Pinsp and Ti in pressure-controlled ventilation modes
Applies for mains operation or in the first 5 minutes of
battery operation with a fully charged battery. In other
situations, the peak inspiratory flow may be restricted
to 75 L/min.
Pressure-related settings
Inspiratory pressure Pinsp
Setting range PEEP +5 to 80 hPa (cmH
2O)
(7 to 80 hPa (cmH
2O) when PEEP = Off)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±3 hPa (cmH
2O) (the greater
value applies)
Pressure limitation Pmax
Setting range PEEP +5 to 80 hPa (cmH
2O)
(7 to 80 hPa (cmH
2O) when PEEP = Off)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±3 hPa (cmH
2O) (the greater
value applies)
Relative pressure support above PEEP
Psupp in CPAP / PSV and PC - SIMV / PS
modes
Setting range Off, 3 to (80 - PEEP) hPa (cmH
2O)
(Off, 3 to 78 hPa (cmH
2O) when PEEP = Off)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±3 hPa (cmH
2O) (the greater
value applies)
Relative pressure support above PEEP
Psupp in VC - SIMV / PS and VC - SIMV /
PS / AutoFlow modes
Setting range Off, 3 to (Pmax - PEEP) hPa (cmH
2O)
(Off, 3 to (Pmax - 2) hPa (cmH
2O) when PEEP = Off)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±3 hPa (cmH
2O) (the greater
value applies)
Positive end-expiratory pressure PEEP
Setting range Off, 2 to 35 hPa (cmH
2O)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±2 hPa (cmH
2O) (the greater
value applies)