Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 279
Technical data
15.8 Breathing system
Inspiratory pressure (Pressure) in the One-
step recruitment maneuver
Setting range PEEP +1 to 80 hPa (cmH
2O)
(3 to 80 hPa (cmH
2O) when PEEP = Off)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±2 hPa (cmH
2O) (the greater
value applies)
Highest inspiratory pressure (Pinsp max) in
the Multi-step recruitment maneuver
Setting range 15 to 80 hPa (cmH
2O)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±2 hPa (cmH
2O) (the greater
value applies)
Highest PEEP (PEEP max) in the Multi-step
recruitment maneuver
Setting range PEEP to 35 hPa (cmH
2O)
(2 to 35 hPa (cmH
2O) when PEEP = Off)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±2 hPa (cmH
2O) (the greater
value applies)
Pressure amplitude above PEEP: ∆Pressure
in the Multi-step recruitment maneuver
Setting range 5 to 30 hPa (cmH
2O)
Increment value 1 hPa (cmH
2O)
Accuracy ±10 % of the set value or ±2 hPa (cmH
2O) (the greater
value applies)
Minimum pressure at patient port in accor-
dance with ISO 80601-2-13
–3 hPa (cmH
2O)
Total volume (when applying a maximum
tidal volume of 1500 mL)
Without CO
2 absorber 2.18 ± 0.2 L
With CLIC absorber 800+ disposable
CO
2 absorber
3.53 ±0.2 L
With reusable CO
2 absorber and
Drägersorb 800+
3.67 ±0.2 L
Information The total volume of the breathing system is typically
smaller and is calculated based on the set tidal volume
plus a reserve volume for compliance compensation.