280 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Technical data
Compliance (without breathing circuit)
In the Man / Spon mode (including
disposable CO
2 absorber)
2.7 ±0.2 mL/hPa (mL/cmH
2O)
corresponds to 81 ±6 mL at 30 hPa (cmH
2O)
for the "Neonates" patient category
2.9 ±0.2 mL/hPa (mL/cmH
2O)
corresponds to 87 ±6 mL at 30 hPa (cmH
2O)
for the "Pediatric patients" patient category
3.0 ±0.2 mL/hPa (mL/cmH
2O)
corresponds to 90 ±6 mL at 30 hPa (cmH
2O)
for the "Adults" patient category
In mechanical ventilation modes 1.0 ±0.2 mL/hPa (mL/cmH
2O)
corresponds to 30 ±6 mL at 30 hPa (cmH
2O)
for the "Neonates" patient category
1.2 ±0.2 mL/hPa (mL/cmH
2O)
corresponds to 36 ±6 mL at 30 hPa (cmH
2O)
for the "Pediatric patients" patient category
1.3 ±0.2 mL/hPa (mL/cmH
2O)
corresponds to 39 ±6 mL at 30 hPa (cmH
2O)
for the "Adults" patient category
Total leakage of the breathing system <150 mL/min at 30 hPa (cmH
2O) standardized to
BTPS conditions
Filling volume of the CO
2 absorber
Reusable CO
2 absorber 1500 mL ±50 mL
Disposable CO
2 absorber CLIC absorber
800+
1300 mL ±50 mL
Disposable CO
2 absorber CLIC absorber
Free
1200 mL ±50 mL
Flexible breathing bag arm
Volume 0.11 L ±50 mL
Compliance 0.11 ±0.05 mL/hPa (mL/cmH
2O)
corresponds to 3.3 ±1.5 mL at 30 hPa (cmH
2O)
Rigid breathing bag arm
Volume 0.13 L ±50 mL
Compliance 0.13 ±0.05 mL/hPa (mL/cmH
2O)
corresponds to 3.9 ±1.5 mL at 30 hPa (cmH
2O)
APL valve
Setting range Open, 5 to 70 hPa (cmH
2O)
Accuracy (at a flow of 20 ±1 L/min): ±20 % of the set value or ±3 hPa (the greater value
applies), but not more than +10 hPa (cmH
2O)
Pressure drop at 30 L/min (ATPD), fully
opened
Dry: 2.1 ±0.3 hPa (cmH
2O)
Wet: 2.2 ±0.3 hPa (cmH
2O)