Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 281
Technical data
15.9 External fresh-gas outlet
15.10 Anesthetic gas receiving system (AGS)
15.11 Active AGS
Recommendation for breathing hoses
Information All compliances and volumes specified include inspira-
tory and expiratory filters.
Neonates (typical VT <50 mL) Maximum compliance: 2.0 mL/hPa (mL/cmH
2O)
Pediatric patients (typical VT between
50 and 300 mL)
Maximum compliance: 4.0 mL/hPa (mL/cmH
2O)
Adults (typical VT >300 mL) Maximum compliance: 6.0 mL/hPa (mL/cmH
2O)
Maximum length 200 cm (78.7 in)
350 cm (137.8 in) (leads to limitations in compliance
compensation and pressure accuracy)
Recommendation for a breathing bag hose
(if no breathing bag arm is being used)
Maximum length 180 cm (70.9 in)
350 cm (137.8 in) with increased fresh-gas consump-
tion
Recommendation for the breathing bag
size
Volume 0.5 L to 5.0 L (at least double the value of the tidal vol-
ume)
Connection 22 mm (outer taper), 15 mm (inner taper), ISO
Delivery See "Fresh-gas delivery"
Pressure limitation Not pressure-limited
Information The device is equipped with an active or a passive
anesthetic gas receiving system.
Information This system is designed for connection to an anes-
thetic gas scavenging system which works with a suc-
tion flow.
Suction flow
Normal range 30 to 50 L/min
At lower end of restricted range ≥10 L/min
Maximum fresh-gas flow to prevent con-
taminating ambient air
For external breathing systems (normal
range)
9L/min