282 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Technical data
15.12 Passive AGS
15.13 Measuring systems and displays
For external breathing systems (restricted
range)
5L/min
For internal breathing systems (restricted
range)
7L/min
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter 3 to 6 mm
Maximum inlet flow 500 mL/min
Information This system is designed for an anesthetic gas scav-
enging system which works without suction flow.
To limit contamination of the ambient air in compliance
with ISO 80601-2-13, follow the specifications for the
scavenging hose.
Scavenging hose
Maximum length 8 m (26 ft)
Minimum diameter 19 mm (0.75 in)
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter 3 to 6 mm
Maximum inlet flow 500 mL/min
Information Alarms and internal calculations are based on values
that are more precise than the displayed values. Con-
sequently, there may be minor deviations between the
current alarm status and the displayed measured val-
ues.
Airway pressure Paw
Plateau pressure Pplat
Positive end-expiratory pressure PEEP
Peak inspiratory pressure PIP
Mean airway pressure Pmean
Range –20 to +99 hPa (cmH
2O)
Accuracy ±4 % of the measured value or ±2 hPa (cmH
2O) (the
greater value applies)
Resolution of displayed value 1 hPa (cmH
2O)
Airway pressure gauge
Range –20 to 80 hPa (cmH
2O)
Accuracy ±5 % of the measured value or ±2 hPa (cmH
2O) (the
greater value applies)
Resolution of displayed value 5 hPa (cmH
2O)