Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 283
Technical data
Volume
Information The measured volume values displayed already take
account of the compliance of the breathing circuit.
For devices with the integrated patient-gas measure-
ment module (PGM), the determined gas composition
and the sample gas flow of the gas measurement are
also taken into consideration as soon as CO
2 respira-
tory phases are detected.
For devices without integrated PGM, the specified
accuracies apply if the flow correction is correctly set
(see the "Flow correction" chapter).
The accuracies specified apply under test conditions in
accordance with ISO 80601-2-13 (ventilation with dry
oxygen, corrected to BTPS conditions).
Expiratory tidal volume VT
Range 0 to 2500 mL
Lowest detectable VT ≤3 mL for the "Pediatric patients" and "Neonates"
patient categories
≤20 mL for the "Adults" patient category
Accuracy ±4 mL in the range below 20 mL
±10 % of the measured value or ±4 mL (the greater
value applies) in the range from 20 to 150 mL
±10 % of the measured value or ±15 mL (the greater
value applies) in the range above 150 mL
Resolution of displayed value 1 mL
Minute volume
Total MV
Mandatory MVmand
Spontaneous MVspon
Range 0.00 to 40.0 L/min
Accuracy ±10 % of the measured value or ±0.15 L/min (the
greater value applies)
Resolution of displayed value 0.01 L/min (for MV <1 L/min) or 0.1 L/min (for MV
≥1L/min)
T
10...90 <45 s (for RR ≥6/min)
<105 s (for RR <6 /min)
Minute volume MV
Range 0.00 to 40.0 L/min
Accuracy ±10 % of the measured value or ±0.15 L/min (the
greater value applies)
Resolution of displayed value 0.01 L/min (for MV <1 L/min) or 0.1 L/min (for MV
≥1L/min)
T
10...90 <45 s (for RR ≥6/min)
<105 s (for RR <6 /min)