284 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Technical data
Respiratory rate
Total RR
Mandatory RRmand
Spontaneous RRspon
Range 0 to 150 /min
Accuracy ±10 % of the measured value or ±1 /min (the greater
value applies)
Resolution of displayed value 1 /min
T
10...90 <45 s (for RR ≥6/min)
<105 s (for RR <6 /min)
Respiratory rate RR
Range 0 to 150 /min
Accuracy ±10 % of the measured value or ±1 /min (the greater
value applies)
Resolution of displayed value 1 /min
T
10...90 <45 s (for RR ≥6/min)
<105 s (for RR <6 /min)
Supply pressures
Central gas supply system (electronic mea-
surement)
Range 0.0 to 10.0 kPa x 100 [bar]
0 to 145 psi
0.00 to 1.00 MPa
Accuracy (up to 7 kPa x 100 [bar]) ±4 % or ±0.2 kPa x 100 [bar] (the greater value applies)
±4 % or ±3 psi (the greater value applies)
±4 % or ±0.02 MPa (the greater value applies)
Resolution of displayed value 0.1 kPa x 100 [bar]
1psi
0.01 MPa
Central supply (physical measurement with
pressure gauge)
Range 0 to 10 kPa x 100 [bar]
0 to 160 psi
0 to 1.0 MPa
Accuracy ±(4 % of the measurement range + 8 % of the dis-
played value) in the range above 1.0 kPa x 100 [bar]
(20 psi or 0.1 MPa)
Resolution of displayed value 0.5 kPa x 100 [bar]
20 psi
0.05 MPa
Gas cylinders (electronic measurement)
Range 0 to 285 kPa x 100 [bar]
0 to 4140 psi
0.0 to 28.5 MPa