Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 285
Technical data
Accuracy ±4 % or ±6 kPa x 100 [bar] (the greater value applies)
±4 % or ±87 psi (the greater value applies)
±4 % or ±0.6 MPa (the greater value applies)
Resolution of displayed value 1 kPa x 100 [bar]
1psi
0.1 MPa
Results in the system test
Total leakage
Range 10 to 5000 mL/min (measured at BTPS, based on a
pressure of 30 hPa)
Accuracy ±25 % of the measured value or ±150 mL/min (the
greater value applies)
Resolution of displayed value 1 mL/min
Leakage during mechanical ventilation
Range 10 to 1000 mL/min (measured at BTPS, based on a
pressure of 30 hPa)
Accuracy ±25 % of the measured value or ±50 mL/min (the
greater value applies)
Resolution of displayed value 1 mL/min
Compliance of the breathing circuit
Range 0.0 to 9.9 mL/hPa (mL/cmH
2O), measured at BTPS
Accuracy ±30 % of the measured value or ±0.2 mL/hPa
(mL/cmH
2O) (the greater value applies)
Resolution of displayed value 0.1 mL/hPa (mL/cmH
2O)
Waveform display
Waveforms O
2 concentration (only with PGM)
Primary anesthetic gas concentration (only with PGM)
CO
2 concentration (only with PGM)
Airway pressure
Volume (only for loops)
Flow
Sweep speed 6.25; 12.5; 25.0 mm/s
Sweep speed 6.25; 12.5 mm/s
Loops Pressure-Volume
Flow-Volume
Scale
Airway pressure –20 to 80 hPa (cmH
2O)
Flow –120 to 120 L/min
Volume 0 to 2000 mL
O
2 0 to 100 Vol%
CO
2 0 to 100 mmHg (0 to 12 Vol%, 0 to 12 kPa)
Halothane 0 to 5 Vol% (kPa)