EasyManua.ls Logo

Dräger Atlan A100

Dräger Atlan A100
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 289
Technical data
CO2
Measuring principle Infrared spectrometry (consumption-free)
Range 0.0 to 13.6 Vol%
0.0 to 13.6 kPa
0 to 102 mmHg
Accuracy ±(0.43 Vol% + 8 % of the measured value)
±(0.43 kPa + 8 % of the measured value)
±(3.3 mmHg + 8 % of the measured value)
Resolution of displayed value 0.1 Vol%
0.1 kPa
1 mmHg
T
10...90 time of the sensor <350 ms
N
2O
Measuring principle Infrared spectrometry (consumption-free)
Range 0 to 100 Vol%
Accuracy ±(2 Vol% + 8 % of the measured value)
Resolution of displayed value 1 Vol%
T
10...90 time of the sensor <500 ms
Anesthetic gases
Measuring principle Infrared spectrometry (consumption-free)
Range
Halothane 0.00 to 8.50 Vol% (kPa)
Isoflurane 0.00 to 8.50 Vol% (kPa)
Enflurane 0.00 to 10.00 Vol% (kPa)
Sevoflurane 0.00 to 10.00 Vol% (kPa)
Desflurane 0.0 to 20.0 Vol% (kPa)
Accuracy ±(0.2 Vol% + 15 % of the measured value)
±(0.2 kPa + 15 % of the measured value)
Resolution of displayed value 0.1 Vol% (kPa) for desflurane
0.01 Vol% (kPa) for all other anesthetic gases
T
10...90 time of the sensor <500 ms
Anesthetic agent identification
Detection Automatic
Primary gas At the latest at 0.3 Vol%
Secondary gas At the latest at 0.4 Vol% or 0.1 xMAC (the greater value
applies)
With a desflurane concentration greater than 4 Vol%,
mixture detection occurs at the latest when the concen-
tration of the second anesthetic gas rises above 10 %
of the desflurane concentration.
The secondary gas becomes the primary gas when the
expiratory xMAC value is more than 0.2 xMAC above
that of the primary gas.

Table of Contents

Other manuals for Dräger Atlan A100

Related product manuals