Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 289
Technical data
CO2
Measuring principle Infrared spectrometry (consumption-free)
Range 0.0 to 13.6 Vol%
0.0 to 13.6 kPa
0 to 102 mmHg
Accuracy ±(0.43 Vol% + 8 % of the measured value)
±(0.43 kPa + 8 % of the measured value)
±(3.3 mmHg + 8 % of the measured value)
Resolution of displayed value 0.1 Vol%
0.1 kPa
1 mmHg
T
10...90 time of the sensor <350 ms
N
2O
Measuring principle Infrared spectrometry (consumption-free)
Range 0 to 100 Vol%
Accuracy ±(2 Vol% + 8 % of the measured value)
Resolution of displayed value 1 Vol%
T
10...90 time of the sensor <500 ms
Anesthetic gases
Measuring principle Infrared spectrometry (consumption-free)
Range
Halothane 0.00 to 8.50 Vol% (kPa)
Isoflurane 0.00 to 8.50 Vol% (kPa)
Enflurane 0.00 to 10.00 Vol% (kPa)
Sevoflurane 0.00 to 10.00 Vol% (kPa)
Desflurane 0.0 to 20.0 Vol% (kPa)
Accuracy ±(0.2 Vol% + 15 % of the measured value)
±(0.2 kPa + 15 % of the measured value)
Resolution of displayed value 0.1 Vol% (kPa) for desflurane
0.01 Vol% (kPa) for all other anesthetic gases
T
10...90 time of the sensor <500 ms
Anesthetic agent identification
Detection Automatic
Primary gas At the latest at 0.3 Vol%
Secondary gas At the latest at 0.4 Vol% or 0.1 xMAC (the greater value
applies)
With a desflurane concentration greater than 4 Vol%,
mixture detection occurs at the latest when the concen-
tration of the second anesthetic gas rises above 10 %
of the desflurane concentration.
The secondary gas becomes the primary gas when the
expiratory xMAC value is more than 0.2 xMAC above
that of the primary gas.