Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 291
Technical data
15.18 Operating characteristics
Sensor sampling rate <40 ms
Typical delay of the measured value <15 s
Mains power supply
Mains voltage 100 to 240 V AC at 50/60 Hz
Maximum power consumption 4 A
Power cable
Maximum length 5 m (16.4 ft)
Protective ground resistance Maximum 0.1 Ω
Operating voltage ≥250 V
Operating current ≥10 A
Current consumption at 230 V AC
When device is switched off with switched-on
main switch and fully charged battery
<0.18 A
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2,
PEEP = 0 hPa, FG flow = 4 L/min O
2) without
charging the internal battery
<0.40 A
Maximum 2 A
Current consumption at 110 V AC
When device is switched off with switched-on
main switch and fully charged battery
<0.38 A
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2,
PEEP = 0 hPa, FG flow = 4 L/min O
2) without
charging the internal battery
<0.84 A
Maximum 4 A
Power consumption
When device is switched off with switched-on
main switch and fully charged battery
<40 W
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2,
PEEP = 0 hPa, FG flow = 4 L/min O
2) without
charging the internal battery
<95 W
Maximum 400 W
Peak inrush current Approx. 8 to 14 A
Approx. 6 to 10 A quasi-RMS
Internal battery
Type Lead battery
Sealed, maintenance-free
Nominal capacity 7.2 Ah