292 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Technical data
Nominal voltage 24 V
Fuse F15A 80V UL248-14, breaking capacity 1000 A, size
19.7 mm x 19 mm x 5 mm
Current Maximum 15 A
Backup time with fully charged battery
During mechanical ventilation (VC - CMV,
VT = 500 mL, RR = 10 /min, I:E = 1:2,
PEEP = 5 hPa, FG flow = 10 L/min O
2) and
with less than 5 completed battery dis-
charge cycles
90 min or longer
Typically 120 min
During mechanical ventilation (VC - CMV,
VT = 500 mL, RR = 10 /min, I:E = 1:2,
PEEP = 5 hPa, FG flow = 10 L/min O
2) and
with less than 20 completed battery dis-
charge cycles
45 min or longer
During ventilation in Man / Spon mode
and with less than 20 completed battery
discharge cycles
90 min or longer
Charging time (to achieve at least 30 minutes
backup time)
At least 8 hours
Charging power Maximum 50 W
Gas supply
Gas quality requirements
Oil content
<0.1 mg/m
3
Dew point 5 °C (41 °F) at ambient temperature
Particle size Dust-free air (filtered with pore size <1 μm)
Supply pressure for O
2, Air, N2O 2.7 to 6.9 kPa x 100 [bar]
39 to 100 psi
0.27 to 0.69 MPa
Maximum short-term peak inlet flow with elec-
tronically controlled gas mixer at 6.9 kPa x 100
[bar] (100 psi or 0.69 MPa) supply pressure
O
2 135 L/min (applies only when there is no distribution
piece for the central O
2 supply)
Air
Without bronchial suction system 50 L/min
Including a directly connected bronchial
suction system
130 L/min
N
2O40L/min
Distribution piece for central O
2 supply
Maximum permissible flow 20 L/min
Distribution piece for central Air supply
Maximum permissible flow 70 L/min
Drive gas Not needed