296 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Technical data
15.20 Relevant standards
Pin 9 Not used
Housing SHLD-GND
USB port Only connect USB mass storage devices that do not
have their own power supply. Do not connect any
charging cables.
Type Connector type A; USB 2.0
Supported devices USB flash drive formatted with FAT16 or FAT32
Network port
Information Only for Dräger Remote Service
Only connect devices or networks that meet the
requirements of IEC 60950-1 / IEC 62368-1 for
ungrounded SELV circuits or the requirements of
IEC 60601-1 (as of the 2nd edition) for touchable sec-
ondary circuits with a maximum nominal voltage of
24 V DC.
Type RJ45 plug
Transfer speed 100Base-TX, IEEE 802.3 Clauses 24 and 25 (requires
at least a CAT5 cable)
10Base-T, IEEE 802.3 Clause 14 (requires at least a
CAT3 cable)
Connector for external workplace light Only for workplace lights approved by Dräger, see list
of accessories.
Information In addition to the standards listed here, this medical
device also meets various other standards.
IEC 60601-1
Medical electrical equipment
Part 1:
General requirements for basic safety and essential
performance
IEC 60601-1-2
Medical electrical equipment
Part 1-2:
General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic
disturbances - Requirements and tests
IEC 60601-1-8
Medical electrical equipment
Part 1-8:
General requirements for basic safety and essential
performance - Collateral standard: Alarm systems -
General requirements, tests and guidance for alarm
systems in medical electrical equipment and medical
electrical systems
ISO 80601-2-13
Medical electrical equipment
Part 2-13:
Particular requirements for basic safety and essential
performance of an anaesthetic workstation