Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 295
Technical data
15.19 Interfaces and ports
Transmitter power ≤42 dBμA/m (200 mW ±1 dB)
Modulation ASK (amplitude shift keying)
Electromagnetic compatibility Tested in compliance with IEC 60601-1-2
Protection classes
Device I, in compliance with IEC 60601-1
Applied parts (connections for breathing
hoses)
TYPE BF
Degree of protection IP11 against vertically falling dripping water under typi-
cal conditions of use in accordance with the intended
use (with 0° inclination of the device)
IP10 at >0° and ≤5° inclination of the device
In accordance with IEC 60529, meets ISO 80601-2-13
Classification Medical Device Europe Class II b
UMDNS code 10-134
Use of latex The device is made without natural rubber latex.
Serial ports COM 1 and COM 2
Information Only connect devices that meet the requirements of
IEC 60950-1 / IEC 62368-1 for ungrounded SELV cir-
cuits or the requirements of IEC 60601-1 (as of the 2nd
edition) for touchable secondary circuits with a maxi-
mum nominal voltage of 24 V DC.
Use COM 2 to connect devices for external patient-gas
measurement with the MEDIBUS communication pro-
tocol.
Protocol MEDIBUS.X
Alarm delay time (measured as of the query
time)
<3 s for baud rates ≥9600
<10 s for baud rates <9600
Connector 9-way Sub-D
Baud rate 1200, 2400, 4800, 9600, 19200, 38400 baud
Data bits 8
Parity Even
Stop bits 1
Pin assignment
Pin 1 Not used
Pin 2 RXD
Pin 3 TXD
Pins 4, 6 Pins 4 and 6 are connected internally
Pin 5 SHLD-GND
Pins 7, 8 Pins 7 and 8 are connected internally