52 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Overview
Backup manual mode with a mechanically controlled gas mixer with
electronic flow measurement
1. Open the flap (1).
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
3. Ventilate the patient manually.
a. During mechanical ventilation or in Man/Spon ventilation mode:
Ventilate manually with the breathing bag.
b. When using the external fresh-gas outlet:
Ventilate manually with the breathing bag on the non-rebreathing system.
4. Set the fresh-gas flow.
5. Check the vaporizer setting.
4067747441