54 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Overview
3.2.12.3 Overview
The following table gives an overview of the integrated safety functions which come
into effect if problems occur during operation:
Fault Safety function
Leakage Pressure regulation and PEEP compensation
In mechanical ventilation modes, the set PEEP is
actively adjusted even in the case of small leaks.
In pressure-controlled modes, the pressure is con-
trolled during the entire respiratory cycle.
Mains power supply
failure
Uninterruptible power supply provided by internal battery
Device runtime in battery operation (see "Technical
data", page 272)
Automatic deactivation of the breathing system warmer
to increase battery runtime
Mains power supply
failure and battery
discharged
Manual ventilation and spontaneous breathing are
available
Emergency O
2 delivery (electronically controlled gas
mixer)
Fresh-gas delivery (mechanically controlled gas mix-
ers)
Delivery of anesthetic agents via connected vaporizers
Failure of the central
gas supply system
Use of the connected gas cylinders
Complete failure of
the gas supply
Mechanical ventilation with ambient air possible (the
hose with the breathing bag will have to be removed for
this)
No anesthetic agent delivery from the connected
vaporizers possible; switch to intravenous anesthetic
agent required
Flow measurement
failure
Mechanical ventilation can be continued.
Restrictions with regard to displayed measured values,
measurement accuracies, and when triggering manda-
tory breaths are possible.
Failure of fresh-gas
delivery (electroni-
cally controlled gas
mixer only)
Emergency O
2 delivery (see "Emergency O2 delivery
(electronically controlled gas mixer)", page 50)
Anesthetic agent delivery from connected vaporizers
possible
All ventilation modes are available.
Alternatively:
Backup manual mode
Ventilator failure Manual ventilation or spontaneous breathing possible
Fresh-gas delivery available
Anesthetic agent delivery from connected vaporizers
possible
Alternatively:
Backup manual mode

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals