94 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Assembly and preparation
4.9.2.2 Filling and mounting
1. Push the absorber insert (5) into the absorber container (6).
CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
Handle the soda lime carefully and do not spill it.
WARNING
Risk of carbon monoxide formation
If soda lime based on potassium hydroxide is used, there is a risk of carbon
monoxide formation.
Do not use soda lime based on potassium hydroxide.
Use a suitable soda lime such as Drägersorb Free.
2. Fill the CO
2 absorber with fresh soda lime to the upper mark.
Recommendation: Use Drägersorb 800 Plus or Drägersorb Free.
WARNING
Risk of hypoventilation and incorrect gas measurement
Reuse of the disposable dust filter can increase the filter resistance and impair the
ventilation function of the device.
If the reusable CO
2 absorber is used, always use a disposable dust filter.
Replace the disposable dust filter with every change of soda lime.
3. If soda lime from third-party manufacturers is used, insert a new disposable dust
filter (7). Only use dust filters from the list of accessories.
Only use undamaged filters.
4. Attach the CO
2 absorber (8) to the breathing system from below. Rotate it in the
direction of the arrow until it reaches the stop.
4112041121
5
6

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals