EasyManuals Logo

Dräger Atlan A100 User Manual

Dräger Atlan A100
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 95
Assembly and preparation
Follow the instructions for use for the particular soda lime.
4.9.3 Water trap
Empty or replace the water trap according to its instructions for use.
4.9.4 Connecting the sample line
WARNING
Risk of underpressure in the lungs
In the following cases, the sample gas flow can immediately cause negative
pressure in the lungs, particularly in neonates or pediatric patients:
Filters or hoses are clogged.
The sample gas is being extracted between the patient and an occluded
component.
Ensure the following when ventilating pediatric patients and neonates:
When using a fine-pored filter close to the patient:
Do not connect the sample line between the tube and the filter. Instead,
connect it directly to the filter or the Y-piece.
When using an HME and a sample line port on a tube adapter close to the
patient:
Use only HMEs that are listed in the list of accessories.
Set the alarm limits for MV low and Paw to suitable values.
Observe the following information: "CO
2 measurement with pediatric patients
and neonates", page 321.
WARNING
Risk due to incorrectly connected sample line
If the sample line is connected to the wrong connectors, e.g., connectors on
infusion pumps, fluids may be drawn in instead of sample gas. As a result, the gas
measurement cannot not display correct values or the device can become
damaged.
When connecting the sample line, take care that it is correctly connected.
WARNING
Risk of incorrect measured gas values
Blocked water traps or blocked sample lines prevent correct gas measurement. As
a result, incorrect measured gas values could be displayed.
Only use Dräger sample lines.

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals