Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 465
System configuration
Alarm volume off Yes
No (default)
Yes – the acoustic alarm signal for any
active alarm is deactivated, the Audio
alarms off message and the symbol
appear in the header bar.
No – the acoustic alarm signals for any
active alarm are not affected when you
press the Code button.
All alarms off On
Off (default)
On the following happens when you select
the Code button.
All audible and optical alarm signals are
deactivated at the Cockpit and at the
ICS.
The message area flashes red with the
message All alarms off. The
symbol appears in the header bar.
The All alarms paused setting is set to
No timeout
if the selection All alarms off reminder
is set to On, an alarm tone sounds every
30 seconds to indicate that the alarms
were disabled.
Selecting the Code button again causes
any existing alarm conditions to be
annunciated immediately.
Off – any existing alarm is still annunciated
optically and acoustically when you select
the Code button.
Selection Available settings Description