System configuration
466 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Alarm setup – configuring M540 settings
The M540 settings page configures certain M540
settings. These setting will automatically update the
M540 settings when the M540 is docked. To access
this page, see page 458.
Selection Available settings Description
Transport alarm
volume
– 50% (default) to 100%
(in increments of 10%)
Determines the speaker volume of the M540 at
the Cockpit while the M540 is undocked.
The Cockpit setting adjusts the Transport
alarm volume setting to match the M540 when
it is undocked.
The Transport alarm volume setting is tied to
the minimum alarm volume setting (see
page 461). If the minimum alarm volume is set
to a higher volume than the selected Transport
alarm volume setting, the Transport alarm
volume setting is adjusted to the higher setting.
If the minimum alarm setting is set to a lower
setting than the current Transport alarm
volume setting, the setting does not change.
Transport pulse
tone volume
–Off (default)
–5%
– 10% to 100%
(in increments of 10%)
Determines the pulse tone volume of the M540
while the device is on transport.
Keep device label – Yes (default)
–No
– Yes – The M540 retains the product label of
the Cockpit when it undocks.
– No – the M540 retains the product label
configured in the M540 wireless menu.