Troubleshooting
518 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
! %0 Unplugged
1)
%0 leads off
1)
The parameter value
is replaced by ***
Lead-off condition
detected due to:
broken
cable(s)
disconnected
ECG lead
wires
loose lead
wire(s)
wrong lead
selected
dried out
electrode gel
ECG cable(s)
disconnected from
the M540.
Replace faulty
cable(s).
Reapply gel on
reusable
electrodes and
reapply them or
replace new
disposable
electrodes.
Select another
ECG lead for
processing.
If monitoring
augmented leads,
verify that the
number of
selected leads in
the ECG setup
page is correct.
Check cable(s)
and connection(s)
Replace cable(s)
if necessary
None
None
None
None
None
None
LA lead off
LL lead off
RA lead off
RL lead off
V lead off
V1 lead off
V2 lead off
V3 lead off
V4 lead off
V5 lead off
V6 lead off
V+ lead off
Parameter value The indicated lead
is no longer
attached to the
patient.
Reattach the
electrode to the
patient.
1)
%0 is a placeholder for the parameter label HR or ECG.
Priority Message Parameter field Problem Solution