Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 519
Troubleshooting
!!! Ventricular
fibrillation
VF The reported
arrhythmia was
detected
Check the patient and
treat if necessary.
Rest ECG messages
None ECG Collecting
waveforms
Parameter value Rest ECG was
initiated
Instruct the patient to
lie still.
None ECG busy Parameter value The central station
is already
processing a
report.
Wait a few minutes
before requesting the
report again.
None ECG cannot
connect
Parameter value Connection to
central station is
not possible.
– Check that the
patient is
admitted at the
central station.
– Check that the
central station
has the Rest ECG
option activated.
None ECG report
complete
Parameter value The Rest ECG
report has been
printed
Informational
message – no action
required.
None %0 comm failure
1)
Parameter value The external
device is not
available.
Check the
configuration at the
central station.
None Sending ECG data Parameter value Informational
message.
Informational
message – no action
required.
1)
%0 is a placeholder for the parameter label HR or ECG.
Priority Message Parameter field Problem Solution
NOTE
RRi and 12-lead ECG monitoring are unavailable
when the M540 is in OR mode and the ECG filter
is set to Monitor.