62
Troubleshooting
Error message or warning Cause Remedy
Transmitter is warming up Fault will resolve itself at the end of
the warming-up time.
Electronic fault Replace transmitter or
call DrägerService.
Transmitter operating time about to
expire
Fit new transmitter, page 64 or
call DrägerService.
Transmitter operating time has expired
and the sensor is no longer working
properly
Fit new transmitter, page 64 or
call DrägerService.
Central unit not connected to
mains
Check installation and correct.
Voltage loss Correct voltage loss.
Transmitter 1
Warming up !
Transmitter 1
Mainten. recommend !
Transmitter 1
Mainten. necessary !
Dräger VarioGard
Emerg. power supply !
If the remedies indicated do not solve the problem, please call DrägerService.
Transmitter 1
Error !