68
Menu mode
Setting the language
Displaying peak values
Setting the language
In measurement mode:
● Press [ u ] key.
● Enter password (for maintenance or configuration).
● Press [ u ] key to change over to menu mode.
Display:
● Use [ q ] or [ l ] key to select the » SET LANGUAGE « option.
● Press [ u ] key to activate the menu option selected.
● Use [ q ] or [ l ] key to select the desired language.
Selection: » deutsch «
» english «
» français «
» español «
● Press [ u ] key to set the language selected. The menu option is
displayed again.
Displaying peak values
The peak value is the maximum value of the gas concentration measured by
the transmitter or by the converter module (in the case of O
2
transmitters, the
smallest measured value).
In measurement mode:
● Press [ u ] key.
● Enter password (for maintenance or configuration).
● Press [ u ] key to change over to menu mode.
Display:
● Use [ q ] or [ l ] key to select the » DISPLAY PEAK VALUES « option.
● Press [ u ] key to display the peak value of the transmitter with the lowest
bus address, e.g.:
● Use [ q ] or [ l ] key to display the peak values of the other transmitters
connected to the system bus, e.g.
● Press [ u ] key to quit the function. The menu option is displayed again.
MENU:
MEASUREMENT
MENU:
SET LANGUAGE
Set language ?
english
MENU:
SET LANGUAGE
MENU:
MEASUREMENT
MENU:
DISPLAY PEAK VALUES
Transmitter 1
20.9 Vol% O2
MENU:
DISPLAY PEAK VALUES
Transmitter 2
18 ppm CO