64
Maintenance
Changing transmitter
Changing transmitter
One transmitter or a succession of transmitters can be replaced by new ones
at the end of their operating time without interrupting operation.
Transmitters or transmitter upper parts must only be replaced by identical
new ones (i.e. with the same part number).
Always use brand-new transmitters or Repair / Replacement Exchange
Parts (REX), see page 81.
Before changing transmitters:
Change over from measurement mode to menu mode and enter
configuration password, see page 68.
Choose » CHANGE SLAVES « option and carry out change-over as
described on page 72.